MEDRAD
®
Salient Operation Manual 47
5.9.3. Autofill Default Options
These options govern the behaviour of the autofill feature on the fill screen. Note that only the autofill volume
can be changed on the fill screen. The shuffle volume and the fill speed will always be set to the values entered
here.
If the autofill volume is changed on the fill screen, the new value will remain in effect until the injector is
powered off and on again, when the default value on this options screen will be reloaded.
5.9.4. Routine Protocol Default Options
These options govern the default behaviour of the routine protocol as follows. Note that all three injection
parameters may be over-ridden by entering a new value on the protocol screen prior to performing the injection.
When the injection completes, the default option values set here will be reloaded.
Autofill Volume: The default
volume to appear on the Autofill
screen
Autofill Shuffle Volume: The
plunger will shuffle back and
forward by this amount at the start
of Auto-filling in order to help
reduce the number of bubbles
created during filling
Autofill Speed
Backspace: Delete last character
Enter new values by pressing on
the numbers
Discard Changes; Return to Options
menu
Accept changes; Return to
Option Menu
Accept changes; Return to Option
Menu
Discard Changes; Return to Options
menu
Enter new values by pressing on the
numbers
Backspace: Delete last character
Routine Protocol Contrast Volume:
The default injection volume to
appear on the Routine Protocol
screen. The setting shown indicates
that All Volume has been selected
Routine Protocol Contrast Flow
Rate: The default injection speed to
appear on the Routine Protocol
screen.
Routine Protocol Pressure Limit:
The default pressure limit to appear
on the Routine Protocol screen.
All Volume; The injector will
simply inject all of the volume in
the syringe