Operation Manual
Table Of Contents
- 1. Quick Start
- 2. Introduction
- 3. Getting Started
- 3.1. Installation
- 3.2. Injector Features
- 3.3. Symbols
- 3.3.1. Symbols used in this manual
- 3.3.2. Symbols used on labelling
- 3.3.3. Handswitch and Footswitch
- 3.3.4. Icons and Buttons used on the Injector Touch Screen Graphical Interface
- 3.3.5. Injector Buttons
- 3.3.6. Remote Control Symbols and Buttons
- 3.3.7. Injector Visual Indicators
- 3.3.8. Icons and Buttons used on the RCU Touch Screen Graphical Interface
- 4. Clinical Information
- 5. Operation Description
- 5.1. Turning the System On for the First Time
- 5.2. Syringe, Quick Fill Tube, Fill Spike, and Connector Tubes
- 5.3. Filling a Syringe Using a Quick Fill Tube (QFT) or Fill Spike on the Injector
- 5.4. [Optional] Off-line Filling
- 5.5. Connector Tube Installation
- 5.6. Dismounting a Syringe
- 5.7. Performing an injection
- 5.8. Programming
- 5.9. Options Menu
- 5.10. [RCU models] Operation Description with Remote Control Unit
- 5.10.1. RCU Features–
- 5.10.2. The Injector and the RCU
- 5.10.3. Turning the System On / Off
- 5.10.4. Injector Controller Application
- 5.10.5. Filling State
- 5.10.6. Routine Protocol
- 5.10.7. Arming
- 5.10.8. Check for Air
- 5.10.9. Starting the Injection
- 5.10.10. Injection in Progress
- 5.10.11. Injection Control Buttons
- 5.10.12. Injection Review
- 5.10.13. Protocol Edit
- 5.10.14. Edit Lock
- 5.10.15. Parameter Value Edit
- 5.10.16. Injecting remaining volume
- 5.10.17. Edit Phase Type
- 5.10.18. Open Protocol
- 5.10.19. Protocol Save
- 5.10.20. Common Injecting Alarms / Warnings
- 5.10.21. RCU Manager
- 5.10.22. RCU Setup
- 5.10.23. History Viewer
- 5.10.24. Protocol Manager
- 5.10.25. eGFR (glomerular filtration rate) Calculator
- 5.11. [Optional] DC200 Scanner Interface
- 5.12. [Optional] Protocol Assistance Tool (PAT)
- 6. Cleaning and Maintenance
- 7. Trouble-shooting
- 8. Specifications
- 8.1. Dimensions
- 8.2. Mechanical
- 8.3. Functional
- 8.4. Controls
- 8.5. Environmental
- 8.6. Electrical
- 8.7. Connectors
- 8.8. IEC 60601-1 Classifications
- 8.9. Ground Continuity
- 8.10. RoHS Statement
- 8.11. IEC60601-1-2:2014 (4th Ed) Compliance
- 8.12. Models
- 8.13. Accessories
- 8.14. Manufacturer’s Software License Information
- 8.15. FCC Statement
- 9. Index
MEDRAD
®
Salient Operation Manual 69
In adults, the best equation for estimating glomerular filtration rate (GFR) from serum creatinine is the isotope
dilution mass spectrometry (IDMS)-traceable Modification of Diet in Renal Disease (MDRD) Study equation.
6
All laboratories should be using creatinine methods calibrated to be IDMS traceable.
This IDMS-traceable MDRD study equation calculator is for use with Scr reported in mg/dL.
GFR (mL/min/1.73 m
2
) = 175 x (Scr)-1.154 x (Age)-0.203 x (0.742 if female) x (1.212 if African descent).
From the Injector Controller screen, press
Enter the patient data and assess the appropriate contrast dose accordingly.
6
Reference: Levey AS, Coresh J, Greene T, Stevens LA, Zhang YL, Hendriksen S, Kusek JW, Van Lente F; Chronic Kidney Disease
Epidemiology Collaboration. Using standardized serum creatinine values in the modification of diet in renal disease study equation for
estimating glomerular filtration rate. Ann Intern Med. 2006 Aug 15;145(4):247-54.
Warning: The recommended contrast volume calculated by the eGFR calculator does not alter
the programmed volume in the current injection protocol. The clinician should use the result, in
combination with clinical experience to decide on the dose and separately program the volume into the
injection protocol.
Warning: The model for the calculation is applicable for adults only. The MDRD calculation is
not recommended for paediatric patients.