Operation Manual

Table Of Contents
MEDRAD
®
Salient Operation Manual 73
5.11.3. Inject or Scan Delay
The injector can start the injection before the scanner (scan delay), or the scanner can start the scan before the
injection (inject delay).
1. Follow the scanner manufacturer’s instructions for enabling ISI.
2. Program the scanner.
3. Program the injection protocol into the injector. Once the protocol has been programmed, press the
ARM button on the injector head or the RCU.
4. Check the fluid pathway for air. Once any air has been purged, press the Check For Air button on the
injector head or the RCU
5. Press the scan/inject delay button.
6. Set the value of the delay from the
keypad or slider. The delay can be
set between 1 99 seconds.
7. Toggle between either
a. Scan Delay
b. Inject Delay
8. Initiate the injection by:
a. Scan Delay - pressing the inject button on the RCU, the injector head or infrared control.
b. Inject Delay - pressing the inject button on the RCU.
9. Press the back button to terminate the procedure with the scanner.
Note: When using an inject delay, injections should be started and stopped on the RCU. Do NOT start and
stop injections at the injector head, or by using the infrared remote control.