Operation Manual
Table Of Contents
- 1. Quick Start
- 2. Introduction
- 3. Getting Started
- 3.1. Installation
- 3.2. Injector Features
- 3.3. Symbols
- 3.3.1. Symbols used in this manual
- 3.3.2. Symbols used on labelling
- 3.3.3. Handswitch and Footswitch
- 3.3.4. Icons and Buttons used on the Injector Touch Screen Graphical Interface
- 3.3.5. Injector Buttons
- 3.3.6. Remote Control Symbols and Buttons
- 3.3.7. Injector Visual Indicators
- 3.3.8. Icons and Buttons used on the RCU Touch Screen Graphical Interface
- 4. Clinical Information
- 5. Operation Description
- 5.1. Turning the System On for the First Time
- 5.2. Syringe, Quick Fill Tube, Fill Spike, and Connector Tubes
- 5.3. Filling a Syringe Using a Quick Fill Tube (QFT) or Fill Spike on the Injector
- 5.4. [Optional] Off-line Filling
- 5.5. Connector Tube Installation
- 5.6. Dismounting a Syringe
- 5.7. Performing an injection
- 5.8. Programming
- 5.9. Options Menu
- 5.10. [RCU models] Operation Description with Remote Control Unit
- 5.10.1. RCU Features–
- 5.10.2. The Injector and the RCU
- 5.10.3. Turning the System On / Off
- 5.10.4. Injector Controller Application
- 5.10.5. Filling State
- 5.10.6. Routine Protocol
- 5.10.7. Arming
- 5.10.8. Check for Air
- 5.10.9. Starting the Injection
- 5.10.10. Injection in Progress
- 5.10.11. Injection Control Buttons
- 5.10.12. Injection Review
- 5.10.13. Protocol Edit
- 5.10.14. Edit Lock
- 5.10.15. Parameter Value Edit
- 5.10.16. Injecting remaining volume
- 5.10.17. Edit Phase Type
- 5.10.18. Open Protocol
- 5.10.19. Protocol Save
- 5.10.20. Common Injecting Alarms / Warnings
- 5.10.21. RCU Manager
- 5.10.22. RCU Setup
- 5.10.23. History Viewer
- 5.10.24. Protocol Manager
- 5.10.25. eGFR (glomerular filtration rate) Calculator
- 5.11. [Optional] DC200 Scanner Interface
- 5.12. [Optional] Protocol Assistance Tool (PAT)
- 6. Cleaning and Maintenance
- 7. Trouble-shooting
- 8. Specifications
- 8.1. Dimensions
- 8.2. Mechanical
- 8.3. Functional
- 8.4. Controls
- 8.5. Environmental
- 8.6. Electrical
- 8.7. Connectors
- 8.8. IEC 60601-1 Classifications
- 8.9. Ground Continuity
- 8.10. RoHS Statement
- 8.11. IEC60601-1-2:2014 (4th Ed) Compliance
- 8.12. Models
- 8.13. Accessories
- 8.14. Manufacturer’s Software License Information
- 8.15. FCC Statement
- 9. Index
MEDRAD
®
Salient Operation Manual 73
5.11.3. Inject or Scan Delay
The injector can start the injection before the scanner (scan delay), or the scanner can start the scan before the
injection (inject delay).
1. Follow the scanner manufacturer’s instructions for enabling ISI.
2. Program the scanner.
3. Program the injection protocol into the injector. Once the protocol has been programmed, press the
ARM button on the injector head or the RCU.
4. Check the fluid pathway for air. Once any air has been purged, press the Check For Air button on the
injector head or the RCU
5. Press the scan/inject delay button.
6. Set the value of the delay from the
keypad or slider. The delay can be
set between 1 – 99 seconds.
7. Toggle between either
a. Scan Delay
b. Inject Delay
8. Initiate the injection by:
a. Scan Delay - pressing the inject button on the RCU, the injector head or infrared control.
b. Inject Delay - pressing the inject button on the RCU.
9. Press the back button to terminate the procedure with the scanner.
Note: When using an inject delay, injections should be started and stopped on the RCU. Do NOT start and
stop injections at the injector head, or by using the infrared remote control.