Operation Manual

Table Of Contents
MEDRAD
®
Salient Operation Manual 83
7.1.1. [RCU Models] Trouble-Shooting
Symptoms Actions
Unit will not connect to injector Select a different Wireless ID on both the RCU and injector system and
try to connect again.
RCU will not turn on Check the power supply is connected to the RCU.
Loading/Saving Protocols Fails Check that a formatted USB memory device is inserted into the RCU.
Wireless connection is Intermittent Ensure that there are no other Salient RCU devices within range using
the same Wireless ID
“System Clock Error” on RCU
bootup.
The rechargeable backup battery for the system clock is flat. Charge
the RCU for several hours before rebooting.
In general, the RCU should be operated while on charger.
7.2. Error Messages
Alarms and error messages are indicated by the status bar at the top of the screen changing colour to red or
yellow. Press on the status bar to view the current alarm condition.
If the error is any other than that described in Section 7.1, remove injector from use and call your authorised
service agent.