Operation Manual
Table Of Contents
- 1. Quick Start
- 2. Introduction
- 3. Getting Started
- 3.1. Installation
- 3.2. Injector Features
- 3.3. Symbols
- 3.3.1. Symbols used in this manual
- 3.3.2. Symbols used on labelling
- 3.3.3. Handswitch and Footswitch
- 3.3.4. Icons and Buttons used on the Injector Touch Screen Graphical Interface
- 3.3.5. Injector Buttons
- 3.3.6. Remote Control Symbols and Buttons
- 3.3.7. Injector Visual Indicators
- 3.3.8. Icons and Buttons used on the RCU Touch Screen Graphical Interface
- 4. Clinical Information
- 5. Operation Description
- 5.1. Turning the System On for the First Time
- 5.2. Syringe, Quick Fill Tube, Fill Spike, and Connector Tubes
- 5.3. Filling a Syringe Using a Quick Fill Tube (QFT) or Fill Spike on the Injector
- 5.4. [Optional] Off-line Filling
- 5.5. Connector Tube Installation
- 5.6. Dismounting a Syringe
- 5.7. Performing an injection
- 5.8. Programming
- 5.9. Options Menu
- 5.10. [RCU models] Operation Description with Remote Control Unit
- 5.10.1. RCU Features–
- 5.10.2. The Injector and the RCU
- 5.10.3. Turning the System On / Off
- 5.10.4. Injector Controller Application
- 5.10.5. Filling State
- 5.10.6. Routine Protocol
- 5.10.7. Arming
- 5.10.8. Check for Air
- 5.10.9. Starting the Injection
- 5.10.10. Injection in Progress
- 5.10.11. Injection Control Buttons
- 5.10.12. Injection Review
- 5.10.13. Protocol Edit
- 5.10.14. Edit Lock
- 5.10.15. Parameter Value Edit
- 5.10.16. Injecting remaining volume
- 5.10.17. Edit Phase Type
- 5.10.18. Open Protocol
- 5.10.19. Protocol Save
- 5.10.20. Common Injecting Alarms / Warnings
- 5.10.21. RCU Manager
- 5.10.22. RCU Setup
- 5.10.23. History Viewer
- 5.10.24. Protocol Manager
- 5.10.25. eGFR (glomerular filtration rate) Calculator
- 5.11. [Optional] DC200 Scanner Interface
- 5.12. [Optional] Protocol Assistance Tool (PAT)
- 6. Cleaning and Maintenance
- 7. Trouble-shooting
- 8. Specifications
- 8.1. Dimensions
- 8.2. Mechanical
- 8.3. Functional
- 8.4. Controls
- 8.5. Environmental
- 8.6. Electrical
- 8.7. Connectors
- 8.8. IEC 60601-1 Classifications
- 8.9. Ground Continuity
- 8.10. RoHS Statement
- 8.11. IEC60601-1-2:2014 (4th Ed) Compliance
- 8.12. Models
- 8.13. Accessories
- 8.14. Manufacturer’s Software License Information
- 8.15. FCC Statement
- 9. Index
MEDRAD
®
Salient Operation Manual 85
8.2. Mechanical
Weight 22.9kg (Single)
25.5 kg (Dual)
Height 1315 mm
Floor Area 500mm x 500mm
No of Wheels 4, locking
Type of Wheel Rubber Castor, ball bearing race
Construction Materials Flame retardant ABS panels, Aluminium, Steel
Operating Noise Level < 95 dBA
8.2.1. [RCU Tablet] Mechanical
Weight 1kg
Dimensions 296(W) x 217(H) x 24.1(D) mm
8.2.2. [DC200] Mechanical
Weight 0.5 kg
Dimensions 120mm x 60mm x 30mm
8.3. Functional
Max Pressure Limit 300 psi
Syringe Mounts Two (contrast and saline) [Dual Only]
One (contrast) [Single Only]
Display pressure accuracy +/- 50psi
(User Selectable) Set Pressure Limit 100 to 300 psi in 1 psi increments
Injected flow rate accuracy +/- 5%
Max Flow Rate 10 ml/s
Programmable Flow Rate User-settable in 0.1 ml/s increments from 0.1 to 10ml/s
Maximum Deliverable Volume 190 ml
Displayed volume accuracy +/- 1% or 0.5ml whichever is greater
Programmable Volume User-settable in 1ml increments from 1 to 190ml
Manual Fill rate 0.1 to 10 ml/s
Auto Fill rate 0.1 to 10 ml/s
Multistage Injection Up to 6 phases
Pause Phase Range 1 to 900 seconds in 0.1 second increments