86 MEDRAD
®
Salient Operation Manual
8.4. Controls
Pedestal Line power ON/OFF, double pole switch
IEC power inlet, double fused
Circuit Breaker 3A max
Injector Head Touchscreen Fill control
Protocol Programming
Injector buttons Power on/off
Arm
Remote control Start/stop injection
Hand-switch or footswitch Start/stop injection
8.5. Environmental
Temperature transport/storage -20°C to +60°C
Temperature operation +10°C to +40°C
Humidity transport/storage 10% to 95% RH, non-condensing
Humidityoperation 20% to 80% RH
Barometric Pressure transport/storage 48kPa to 110kPa
Barometric Pressure - operation 70kPa to 106kPa
8.5.1. [RCU Tablet] Environmental
Temperature storage and transportation
-20°C to 50°C
Relative humidity storage and transportation
20% to 93% (40°C)
Atmospheric pressure 86kPa to 106kPa
Temperature - operation
0°C to 35°C
Relative humidity operation 35% to 80%
8.5.2. [DC200] Environmental
Temperature transport/storage -20°C to +60°C
Temperature operation +10°C to +40°C
Humidity transport/storage 10% to 95% RH, non-condensing
Humidity operation 20% to 80% RH
Barometric Pressure transport/storage 48kPa to 110kPa
Barometric Pressure - operation 70kPa to 106kPa