Operation Manual
Table Of Contents
- 1. Quick Start
- 2. Introduction
- 3. Getting Started
- 3.1. Installation
- 3.2. Injector Features
- 3.3. Symbols
- 3.3.1. Symbols used in this manual
- 3.3.2. Symbols used on labelling
- 3.3.3. Handswitch and Footswitch
- 3.3.4. Icons and Buttons used on the Injector Touch Screen Graphical Interface
- 3.3.5. Injector Buttons
- 3.3.6. Remote Control Symbols and Buttons
- 3.3.7. Injector Visual Indicators
- 3.3.8. Icons and Buttons used on the RCU Touch Screen Graphical Interface
- 4. Clinical Information
- 5. Operation Description
- 5.1. Turning the System On for the First Time
- 5.2. Syringe, Quick Fill Tube, Fill Spike, and Connector Tubes
- 5.3. Filling a Syringe Using a Quick Fill Tube (QFT) or Fill Spike on the Injector
- 5.4. [Optional] Off-line Filling
- 5.5. Connector Tube Installation
- 5.6. Dismounting a Syringe
- 5.7. Performing an injection
- 5.8. Programming
- 5.9. Options Menu
- 5.10. [RCU models] Operation Description with Remote Control Unit
- 5.10.1. RCU Features–
- 5.10.2. The Injector and the RCU
- 5.10.3. Turning the System On / Off
- 5.10.4. Injector Controller Application
- 5.10.5. Filling State
- 5.10.6. Routine Protocol
- 5.10.7. Arming
- 5.10.8. Check for Air
- 5.10.9. Starting the Injection
- 5.10.10. Injection in Progress
- 5.10.11. Injection Control Buttons
- 5.10.12. Injection Review
- 5.10.13. Protocol Edit
- 5.10.14. Edit Lock
- 5.10.15. Parameter Value Edit
- 5.10.16. Injecting remaining volume
- 5.10.17. Edit Phase Type
- 5.10.18. Open Protocol
- 5.10.19. Protocol Save
- 5.10.20. Common Injecting Alarms / Warnings
- 5.10.21. RCU Manager
- 5.10.22. RCU Setup
- 5.10.23. History Viewer
- 5.10.24. Protocol Manager
- 5.10.25. eGFR (glomerular filtration rate) Calculator
- 5.11. [Optional] DC200 Scanner Interface
- 5.12. [Optional] Protocol Assistance Tool (PAT)
- 6. Cleaning and Maintenance
- 7. Trouble-shooting
- 8. Specifications
- 8.1. Dimensions
- 8.2. Mechanical
- 8.3. Functional
- 8.4. Controls
- 8.5. Environmental
- 8.6. Electrical
- 8.7. Connectors
- 8.8. IEC 60601-1 Classifications
- 8.9. Ground Continuity
- 8.10. RoHS Statement
- 8.11. IEC60601-1-2:2014 (4th Ed) Compliance
- 8.12. Models
- 8.13. Accessories
- 8.14. Manufacturer’s Software License Information
- 8.15. FCC Statement
- 9. Index
MEDRAD
®
Salient Operation Manual 87
8.6. Electrical
Line Voltage 100-230 V~ ±10%
Line frequency 50/60 Hz
Phase Single
Charging AC only
Operating Batteries or AC
Max Current Consumption, Operating 2.0A peak
Max Current Consumption, Standby 0.5A or less
Consumption, Charging from mains 1A maximum
Current Surge (inlet fuse) Rating 2A maximum
Battery Voltage 12VDC
Battery type Imaxeon 12 V7.2Ah Lead Acid
Battery Hazardous Material Content Lead: 48~53 wt%
Lead Oxide: 23~26%
Lead Sulphate: < 1. wt%
Electrolyte – Sulphuric Acid: 7~10 wt%
Contains no mercury or mercury compounds.
No. of Batteries 2
Recharge time (for 25 injections) 8 hours
Battery Service Life 2 years
Contrast Heat Maintainer operation Auto detects heat maintainer plugged in.
Contrast Heat Maintainer – rating 5W
Contrast Heat Maintainer – temperature range 37±4°C
Category AP/APG Not Applicable
8.6.1. [RCU Tablet] Electrical
AC Adapter Wide mode AC power supply
100 – 240VAC 1.5A 50 – 60 Hz
19V 65W DC to tablet
Wireless IEEE 802.11 b/g/n
Injector module
Frequency Range: 2412.0 – 2462.0 MHz
Output Power: 0.00865 W
RCU Tablet
Frequency Range: 2412.0 – 2462.0 MHz