Operators Manual (purchased after 10 May 2012) User Manual
800-446-7488 35 www.in-situ.com
Alarms Setpoints and Deadbands
High Alarm Mode
l The relay is turned on if the current reading is equal to or greater than the setpoint
value.
l The relay is turned off if the current reading is less than or equal to the setpoint
minus the deadband.
Figure 8.1 High Alarm relay behavior
Low Alarm Mode
l The relay is turned on if the current reading is equal to or less than the setpoint
value.
l The relay is turned off if the current reading is greater than or equal to the setpoint
plus the deadband.
Figure 8.2 Low Alarm relay behavior
Hold (Pause) the Outputs
You may temporarily pause the outputs.
1. From the main screen, select Menu, Hold Outputs. All measurement updates will
stop and the current values and outputs will be held until you select Resume.
2. You can use the View function and arrow keys to scroll through the parameters.