Field Deployment
Program the Instrument
In order to set up a log or download data, you must connect the instrument to a
computer running Win-Situ 5 Software, or a RuggedReader Handheld PC running Win-
Situ Mobile. See page 50.
Position the Instrument
Place the instrument at the desired depth. Position the instrument below the lowest
anticipated water level, but not so low that the pressure sensor range might be
exceeded at the highest anticipated water level. Refer to the tables below to determine
usable depth.
A BaroTROLL Instrument can be deployed with a non-vented
instrument to compensate level data for changes in atmospheric
pressure. Make sure the clocks on both instruments are
synchronized, and install the BaroTROLL in a location that will
never be submerged in water. See page 39.
Range Effective Range Usable Depth
PSIA PSIA kPA Meters Feet
30 15.5 106.9 11 35
100 85.5 589.5 60 197
300 285.5 1968 200 658
500 485.5 3347 341 1120
1000 985.5 6795 693 2273
* Effective range for psia sensors is limited by an estimated
14.5 PSI atmospheric pressure at sea level.
Non-Vented Level TROLL Instrument