User Manual

53
© InnoCare Optoelectronics Corporaon │ X-Ray Products
4.4
Guidelines for Pediatric Applications
InnoCare typically conducts radiologists(Government-certified) usability
studies that compare Yushan detector models to the marketed devices.
Participate radiologists have demonstrated that the images acquired using the
Yushan detectors are deemed to be of diagnostic capability.
Additionally, please review the following scientific literatures regarding
pediatric (https://www.ncbi.nlm.nih.gov/pubmed/29064378) and neonates
(https://www.ncbi.nlm.nih.gov/pubmed/23086629). The verification of Yushan
detectors are using even lower energy or x-ray voltages to acquire the image,
which prove the safety and effectiveness for pediatric and neonates
applications. Please refer to the following paragraphs for detailed explaination.
For pediatric, this study has evaluated the physical image quality
characteristics of five detectors (2 for CR and 3 for FPD) for paediatric
applications, using the IEC RQA5 energy. The results show a clear evolution in
detector performance for both the CR and FPD detectors, with physical image
quality continuing to improve. In this artical, the DQE was evaluated at the
normal operating detector air kerma (DAK) level, defined at 2.5 µGy for all
detectors. The performance of MTF and DQE for Yushan detectors were all
tested and verified at the level of 2.0 µGy, and the test results shows highly
sensitivity for image quality.
For neonates, in this study, they investigated the digital characteristics of the
digital mobile X-ray imaging system equipped with a new FPD for different
tube voltages, and showed that it has an excellent linearity of the digital
characteristics and constant contrast characteristics for tube voltages in the
range of 50-110 kV. The system can be used for significantly reducing the ESD
to neonatal patients (examined by chest X-ray in the NICU) as well as the
X-ray system operator exposure more effectively than is available with
conventional systems. The physical characterization of the Yushan detectors
were obtained with the standard beam conditions RQA5 (in accordance with
IEC 61267:1994) with the x-ray tube voltage 70kV.
Furthermore, please review the following link and reduce pediatric technique
factors accordingly: http://www.imagegently.org
As a general rule, next recommendations shall be observed in pediatrics:
X-Ray Generator must have short exposures times.