Instruction Manual

IDS 430/433 Users Manual Version 1.0
- 13 -
Setup Guide
Initialize to Factory Defaults
Setup the IDS 433/430 in the following order:
The IDS 433/430 is initialized at the factory. The IDS 433/430 should be
initialized when first powered up to guarantee no corruption occurred during
shipping. It must be re-initialized if the program memory (EPROM) or the
parameter memory (EAROM) is replaced. The Initialize function clears the
memory and sets the configuration parameters to their default values. Use
parameter function 70 described in Parameter Functions section (Page 7-7) to
initialize the IDS 433/430.
NOTE: Use the code 10809 for a factory default initialization. Calibration
and setup must be performed after the IDS 433/430 is initialized with this
code.
Setup & "Quick" Calibration Guide
The IDS 433/430 should be configured after initializing and BEFORE calibration.
Set Operation Parameters details all of the scale parameters that are available.
To quickly get the IDS 433/430 up and running, the following Quick Calibration
Guide lists only the mandatory parameters that must be defined. Enter the
following scale parameters using the selection outlined in Set Operation
Parameters and Parameter Functions.
Parameter (section) Purpose (Factory Setting)
1 (Set Operation Parameters) Load Cell mv/V (3)
11 (Set Operation Parameters) Decimal Point (none)
12 (Set Operation Parameters) Count-by (1)
17 (Set Operation Parameters) Full Scale Graduations
(10,000)
18 (Set Operation Parameters) Overload Graduations
(10,200)
60 (Parameter Functions) Calibrate Scale
Calibrate the Weight Indicator
Use Parameter Function 60 to calibrate the weight indicator (page 7-2).