Instruction Manual

IDS 430/433 Users Manual Version 1.0
- 27 -
Set Operation Parameters
This section describes the operation parameters of the IDS 430/433 and how
their settings affect the operation of the weight indicator.
Access Operation Parameters and Parameter Functions
The parameters are accessed by holding the CLEAR key down and then
pressing the ENTER key. The numeric display will prompt “CFG xx” where xx is
the currently selected parameter. Enter the parameter number to be modified
and then press the ENTER key. NOTE: the UNITS key () and the PRINT key
() can be used to scroll the CFG parameter number. Press the UNITS key to
increment the CFG number, press the PRINT key to decrement the CFG
number.
To exit configuration and store changes, press the CLEAR key when “CFG xx” is
being displayed.
To exit configuration and NOT store changes, enter 999 when “CFG xx” is being
displayed. The display prompts “Abort”. Press the ENTER key to abort (exit
without saving changes) or press the CLEAR key if you do not want to abort.
Operation Parameters Reference Tables
The reference tables provide an overview of the operation parameters. The
sections following the reference tables describe the setup parameters in detail
and the procedures for modifying them. Use the ‘Field Settings’ column of the
reference tables to document any changes to the setup parameters.
Reference Table column headings descriptions:
No.: Use this number to access the parameter to be modified. For example to
modify the Digital Filter enter CFG 2.
Parameter Name: The name of the parameter.
Default: The value set at the factory. The Parameter Function “Initialize
Operation Parameters to Factory Defaults” sets the parameter to this value.
Max Value: The maximum value for numeric parameters, the maximum
number of characters for alphanumeric parameters.
Pass: The IDS 430/433 has 4 password levels. Level 1 requires entering
password-1 before accessing the parameter. Level 2 requires password-2.
Level 3 requires password-2 and the Lock Switch to be in the unlocked
position. Level 4 inhibits Scale Basic functions from being uploaded to a PC.
Field Setting: Record any changes to the parameters in the Field Setting
column.