Parameter Functions
The parameter functions are used to set parameters that are too complex for a
single numeric entry. The following table lists the parameter functions:
Parameter Functions
Fn. No Function Name Password Level
59 Display Calibration Audit Number 0
60 Calibrate Scale - deadload first 3
61 Calibrate Scale - span first 3
62 Calibrate Deadload Only 3
63 Adjust Gain Calibration 3
64 Configure Passwords 2
65 Configure Print Formats 2
66 Set Time and Date 2
67 Display Operation Parameters 0
68 Print Operation Parameters 0
69 Diagnostic Tests 0
70 Initialize Operation Parameters to factory defaults. 3
71 Configure Event Monitor 2
72 Function Editor 2&3
73 Initialize Function Memory 2
74 Debug mode for Event monitors and Scale basic fns. 2
75 Enable Peak Detect (optional) 2
76 Set debug mode for event monitors and scale basic fns.
2
78 Erase Id Memory 1
79 Enable Multi-point Linearization (default = disabled[0]) 3
80 Calibrate using Multi-point Linearization 3
81 Analog output offset (default = 0) 1
82 Analog output span (default = 10,000) 1
83 Analog output register (default = 0 [disabled]) 1
The Password Level is used to prevent unauthorized access to parameter
functions. See Parameter Function 64, “Configure Passwords”, for more
information on passwords.
Access Parameter Functions
Hold the CLEAR key down and press the ENTER key. The numeric display
prompts “CFG xx” where xx is the currently selected parameter. Enter a
parameter function number and then press the ENTER key. NOTE: use the
UNITS key () and the PRINT key () to scroll the CFG parameter number.
Press the UNITS key to increment the CFG number, press the PRINT key to
decrement the CFG number.