Instruction Manual

IDS 430/433 Users Manual Version 1.0
- 44 -
Function 62 Calibrate Deadload Only
Use function 62 to re-calibrate the deadload weight without affecting the span
calibration.
1. Hold the CLEAR key down and press the ENTER key to enter the configure
mode.
2. Enter CFG 62. The IDS 430/433 prompts “dEAdld”. Remove all weight from
the scale. Press the ENTER key. The IDS 430/433 reads the scale’s
deadload weight.
3. The IDS 430/433 prompts “CFG 62”. Press the CLEAR key to exit the
configure mode.
Function 63 Adjust Gain Calibration
Use function 63 to make small adjustments to the span calibration factor. Zero
the scale and then place a known weight on the scale.
1. Hold the CLEAR key down and press the ENTER key to enter the configure
mode.
2. Enter CFG 63. The IDS 430/433 displays the weight on the scale along with a
center of digit indication in the least significant digit position of the display (the
minus sign (-) indicates the center of digit.
3. Press the ‘1’ key to adjust by 1, press the ‘2’ key to adjust by 10, the ‘3’ key
to adjust by 100, or the 4 key to adjust by 1000. Use the UNITS key () and
the PRINT key () to increase or decrease the displayed weight until the
displayed weight is the same as the weight on the scale. The center of digit
indicator (-) should be displayed for maximum accuracy.
4. Press the CLEAR key to exit Adjust Gain procedure.
Adjust Gain – 5 Point Linearization
If 5 point Linearization is enabled, the gain adjust procedure begins with the
display alternating between the calibration number (example: Cal 1) and the
calibration weight.
Press the ENTER key to adjust the gain at the displayed calibration point.
Press the UNITS key () to adjust the next calibration point
Press the PRINT key () adjust the previous calibration point.
Press the ZERO key to clear a calibration point. (Must re-calibrate to reuse).