Instruction Manual

IDS 430/433 Users Manual Version 1.0
- 53 -
Function 71 Configure Event Monitor
Use function 71 to enter setpoint data into the Event Monitor. See the “Scale
Basic / Event Monitor” manual” for more information on the event monitor
parameters.
1. Hold the CLEAR key down and press the ENTER key to enter the configure
mode.
2. Enter CFG 71. The IDS 430/433 prompts “Eno”. Enter the Event number to
be modified.
3. The IDS 430/433 prompts “P0 xx” where P0 is parameter 0 and xx is the
current value of parameter 0. Enter new data for P0 or press the ENTER key
to use the displayed data.
4. Use the UNITS key () or the PRINT key () to scroll through the parameter
numbers. Press the CLEAR key to exit from the current Event number. Press
the CLEAR key again to exit from event monitor data entry.
Function 72 Scale Basic Function Editor
Use function 72 to create or modify Scale Basic functions. See the “Scale Basic /
Event Monitor” manual for more information on Scale Basic functions.
1. Hold the CLEAR key down and press the ENTER key to enter the configure
mode.
2. Enter CFG 72. The IDS 430/433 prompts “Fn”. Enter the function number to
be modified.
3. The IDS 430/433 prompts “0- xxx” where 0- is entry zero, and xxx is the
current value for entry zero. Enter new data for 0- or press the ENTER key to
use the displayed data.
4. Use the UNITS key () or the PRINT key () to scroll through the entry
numbers. Press the CLEAR key to exit from the Function Editor.