User Manual

98 ISP-074-575-P1D
98
Inlet filter: Replaceable Teflon/Polypropylene, 1 m
Inlet flow rate: Greater than 300 mL/min
Operating
temperature
range: 0 to 50°C (32 to 122°F)
I/S Certified
Temperature
Range: 0 to 40
0
C (32 to 105
0
F)
Operating 0 to 95% relative humidity
humidity range: (non-condensing)
Operating 0 to 100% relative humidity
humidity range (non-condensing)
with optional
humidity pre-
filter tube
Operating
altitude
range: Up to 10,000 ft (3000 m)
Storage
temperature
range: -10 to 65°C (14 to 149°F)
Storage
humidity range: 0 to 100% relative humidity (non-condensing)
Storage
altitude
range: Up to 15,000 ft (4570 m)
Operating
concentration
range: 0.1 ppm to 10,000 ppm, isobutylene
Accuracy: +/-10% or +/-2 ppm, whichever is greater
Precision: 1% of calibration (calibrated with 100 ppm
isobutylene)
Response time: Less than 3 seconds to 90%
Detection limit: 0.1 ppm isobutylene
Languages: English (standard), can add either Spanish,
French, German or Italian as an option
NOTE: Specifications subject to change without notice.