10 Verify from the Command Prompt window that the PC is receiving data packets
from the instrument.
NOTE: If the ping command prompt displays Request Timed Out or Detection Limit
Unreachable, then connect to the static IP address (192.168.1.99) by following the
procedure in 3000 Micro GC Guide 2, Connecting to 3000 Micro GC via the Static IP
Address.
If connecting to the static IP address of the instrument is unsuccessful, contact
the network administrator to check whether there is a conflict of the newly assigned
IP address or any possible network issue between the instrument and the PC.
Once the network issue is resolved, perform steps in this section again to configure
the instrument to the new IP address.
If connecting to the static IP address is unsuccessful, contact INFICON.
Table 3-1 Changing the Instrument IP Address (continued)
Step Description