User Manual
10
Specifications
Usage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Indoor or Outdoor
Minimum sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . High Sensitivity
0.25 oz./yr. (7 g/a)
Operating temperature range. . . . . . . . . . . . . . . . . . -20 to +50 °C (-4 to 122 °F)
1
Storage temperature range . . . . . . . . . . . . . . -20 °C to +60 °C (-4 °F to + 140 °F)
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95% RH NC Max.
Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000 m (6500 ft.)
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Two “D” cell alkaline batteries
Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Approximately 16 hours
Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Overvoltage category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Weight (with batteries) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.58 kg (1.28 lb.)
1
May be operated for a limited time in lower temperature environments.
Specification Table in Accordance with EN 14624
Minimum sensitivity to R134a, fixed (static) 2 g/yr
Maximum sensitivity to R134a, fixed (static) >50 g/yr
Minimum sensitivity to R134a, moving (dynamic) 2 g/yr
Maximum sensitivity to R134a, moving (dynamic) >50 g/yr
Minimum response/detection time <1 second
Zeroing time 5-7 seconds
Recovery time for 50 g/yr exposure* 12 second
Minimum sensitivity in contaminated environment 2 g/yr
Calibration frequency: Check annually with calibrated leak standard.
*Upper leak detection limit is not specified by INFICON as there is no upper limit to the size
of the leak the detector is able to detect. As no 50 g/yr leak standard was available during
testing, a 31 g/yr leak was substituted.