User Manual
9
Return Authorization Procedure
All defective TEK-Mates should be returned to your wholesaler for warranty
evaluation. If you have any questions, please contact INFICON at 800-344-3304.
NOTE: Do not return your defective unit directly to the factory without first
contacting your wholesaler.
4. False alarms -
TEK-Mate alarms
when the probe is
moved or bumped.
4a. Sensor leads are
bent.
4a. Remove the sensor and
inspect the leads. Straighten
the leads with needle nose
pliers, if necessary, and
reinstall the sensor.
4b. Moisture was
absorbed by the
sensor during a long
period without use.
4b. Run TEK-Mate for at least
20 minutes. The absorption of
moisture does not affect the
life or sensitivity of the sensor.
PROBLEM CAUSE REMEDY