Manual

12
tira93e1 (2014-04) BPG402-SE.cp
The manufacturer-specific profile objects contain the manufacturer's model number
and device configuration data, status and diagnostic data. The objects are
described in the following tables.
Index SI DataType NV Access PM Name
2800 0x02 REAL RW Safe Value Module 1
Safe Value: The value to be used for Safe State = Safe Value.
Index SI DataType NV Access PM Name
2801 0x02 REAL RW Safe Value Module 2
Safe Value: The value to be used for Safe State = Safe Value.
Index SI DataType NV Access PM Name
6000 0x0E BOOL RO tx TxPdoState Module 1
0x11 REAL RO tx Sensor Value Module 1
Is set if the device is not in Safe State (value (I 0x6nn0, SI 0x11) = valid)
TxPdoState Module 1
0
1
Invalid
Valid
The corrected, converted, calibrated final analog input value of the sensor.
Index SI DataType NV Access PM Name
6003 0x01 BOOL RO tx Reading Valid Module 1
0x02 BOOL RO tx Overrange Exceeded
Module 1
0x03 BOOL RO tx Underrange Exceeded
Module 1
Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.
Reading Valid Module 1
0 Invalid
1 Valid
4.3 Manufacturer-specific
Profile Objects
(0x2000...0x5FFF)
4.3.1 Manufacturer
Configuration Module 1
Subindex 0x02
4.3.2 Manufacturer
Configuration Module 2
Subindex 0x02
4.4 Input Area
(0x6000…0x6FFF)
4.4.1 Input Common
Subindex 0x0E
Subindex 0x11
4.4.2 Input Heat Transfer
Subindex 0x01