User guide

5 - 34
PN 074-566-P1B
Composer Elite Operating Manual
12 Heater 2 Current -
Expectations Not Met
0x0000 0800 SC SS Self Clear ACTION(SC): SC shuts off
Heater.
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
NOTE: Does not meet
expectation based on PWM duty
11 Heater 1 Setpoint -
Bad Parameter
0x0000 0400 SC SS Cleared on
read
NOTE: Attempt made to set
parameter above operating
range (>65°C).
10 Heater 2 Setpoint -
Bad Parameter
0x0000 0200 SC SS Cleared on
read
NOTE: Attempt made to set
parameter above operating
range (>68°C).
Table 5-23 System and hardware warnings
Warning
Number
Description Warning Code Set by
CC
1
or SC
2
Sensor
Specific
(SS
3
)
Set/Clear
Trigger
General Notes/Actions
32 Composer Elite
Temperature Outside
Operating Range
0x8000 0000 CC Self clear
31 DeviceNet Installed
but Not Operating
0x4000 0000 CC Self clear NOTE: Indicates that the
DeviceNet module was found
but DeviceNet system is not
currently operating.
30 DeviceNet Failed to
Transmit a Response
0x2000 0000 CC Cleared on
read
NOTE: Failed to transfer
response from CC to ABIC
4
module.
29 Parameters Reset 0x1000 0000 CC Cleared on
read
NOTE: Could be result of
automatic parameter reset due
to compatibility error or from a
manual reset.
28 CC Firmware
Changed
0x0800 0000 CC Cleared on
read
27 Splash Image
Incompatible or Invalid
0x0400 0000 CC Cleared on
read
NOTE: The splash image read is
not compatible with the system
expectations and thus cannot be
displayed.
26 Parameters Upgraded 0x0200 0000 CC Cleared on
read
1
CC = Controller Card,
2
SC = Sensor Card,
3
SS = Specific Sensor,
4
ABIC = Anybus Integrated Circuit,
5
R2= Lorentzian Coefficient of Determination
Table 5-22 System and hardware errors (continued)
Error
Number
Description Error Code Set by
CC1
or SC2
Sensor
Specific
(SS3)
Set/Clear
Trigger
General Notes/Actions
1
CC = Controller Card,
2
SC = Sensor Card,
3