User guide

5 - 35
PN 074-566-P1B
Composer Elite Operating Manual
19 All CC/SC Command
Retries Have Failed
0x0004 0000 CC SS Cleared on
read
NOTE: CC has transmitted a
command to the SC the
maximum number of times
without receiving a response.
Attempt has been abandoned.
18 Sensor Offline 0x0002 0000 CC SS Self Clear
17 Sensor firmware
Changed
0x0001 0000 CC SS Cleared on
read
15 No Signal (Cable
detected)
0x0000 4000 SC SS Self Clear NOTE: No/very low (below
operating threshold) audio
signal. Sensor card will
continually be in
Search mode.
14 Concentration Result
Error
0x0000 2000 SC SS Self Clear NOTE: Generally indicates a
change in gas flow/mixture. SC
will perform new sweep if it
persists more than
a few seconds.
13 Bad peak Shape 0x0000 1000 SC SS Self Clear NOTE: Indicated by low
R2
5
/ high Std Err.
Table 5-23 System and hardware warnings (continued)
Warning
Number
Description Warning Code Set by
CC
1
or SC
2
Sensor
Specific
(SS
3
)
Set/Clear
Trigger
General Notes/Actions
1
CC = Controller Card,
2
SC = Sensor Card,
3
SS = Specific Sensor,
4
ABIC = Anybus Integrated Circuit,
5
R2= Lorentzian Coefficient of Determination