User guide

6 - 6
PN 074-566-P1B
Composer Elite Operating Manual
Figure 6-5 Data storage
In the Save Options pane, the Interval number allows you to skip data points and
only save data based on the entered Interval number. Acceptable Interval values
are 1 - 99. Composer Elite takes 1 measurement per second, so if the Interval is
set to 1 (default value), every measurement is recorded to the log file. Setting
Interval to 2 would log data every 2 seconds, 3 for every 3 seconds, etc. This is
particularly useful when logging data over a long period of time.
The check boxes in the Save Options pane select the data of interest to be saved
in the datalog. By default, all data is selected.
If the drive runs out of space, datalogging will
discontinue and no additional data will be stored.
6.3.4 Diagnostic Files
Composer Elite Multi Sensor Software automatically creates diagnostic files which
monitor instrument performance. These files can be very valuable in identifying and
troubleshooting any potential problems. There are two file types and as many as
six files are saved to the root of the My Documents folder. All files older than seven
days are automatically deleted. The following file types and contents are created: