User guide

1 - 13
PN 074-566-P1B
Composer Elite Operating Manual
1.5.6 Computer Requirements for Composer Elite Multi-Sensor Software
Processor . . . . . . . . . . . . . . . . . . . . . 1 GHz
Memory . . . . . . . . . . . . . . . . . . . . . . 512 MB RAM
Display Resolution . . . . . . . . . . . . . . 1280 x 1024
Storage . . . . . . . . . . . . . . . . . . . . . . . 5 MB of Hard Drive Space
Operating System. . . . . . . . . . . . . . . Windows XP (SP3), Windows 7
Communications. . . . . . . . . . . . . . . . RS-232 Serial Port, or USB-Serial Adapter,
or DeviceNet Card
Other . . . . . . . . . . . . . . . . . . . . . . . . Microsoft
®
.Net Framework 4 Client Profile
(x86 and x64)
1.5.7 DeviceNet Requirements for Composer Elite Multi-Sensor Software
Interface Card. . . . . . . . . . . . . . . . . . Beckhoff FC5201 DeviceNet Interface Card
Connectivity . . . . . . . . . . . . . . . . . . . 1-Channel, PCI bus
Software . . . . . . . . . . . . . . . . . . . . . . Beckhoff TwinCAT I/O
Power . . . . . . . . . . . . . . . . . . . . . . . . 24 V(dc), external power supply
NOTE: These requirements do not apply when DeviceNet is used in conjunction
with other software.
1.5.8 Mounting Requirements
The Sensor Control Unit is designed for bench-top operation or mounting in a
standard 48.26 (19 in.) rack using the optional Full Rack Extender
(PN 782-900-007). The Sensor Control Unit occupies half the width of a
standard rack, so a blank-off is provided with the optional Full Rack Extender.
The Acoustic Sensor must be attached to a rigid mounting surface using the
M4 threaded holes in the Acoustic Sensor cover. Maximum depth of screw
penetration into M4 mounting holes is 6.4 mm (0.25 in.).
Dimension from VCR-4 gland-to-gland is 12.4 cm (4.88 in.).
1.5.9 Ventilation Requirements
The Sensor Control Unit is adequately ventilated when operating on a bench
top or when installed in a standard 48.26 cm (19 in.) rack using the optional
Full Rack Extender (PN 782-900-007).
The Acoustic Sensor does not require ventilation.