User guide

3 - 5
PN 074-566-P1B
Composer Elite Operating Manual
3.3.2 Composer Elite Screen Descriptions
Boot Screen, see section 3.3.2.1
Indicators Displayed On Screens, see section 3.3.2.2 on page 3-6
Overview Screen, see section 3.3.2.3 on page 3-6
System Overview Screen, see section 3.3.2.4 on page 3-8
Errors and Warnings Screen, see section 3.3.2.5 on page 3-9
Sensor Screen, see section 3.3.2.6 on page 3-10
Sensor Parameters Screen, see section 3.3.2.7 on page 3-12
3.3.2.1 Boot Screen
During power-up of the Sensor Control Unit, the INFICON logo is displayed. The
following information is briefly displayed before the System Overview screen is
displayed:
Version x.xx.xx HW x (Control card firmware version and hardware version)
Loading....... Please Wait (while Data, Sensor, and DeviceNet statuses are
being determined) followed by Loading....... Complete
Data Wait followed by:
Data OK (if parameter values were retained in memory)
Data Reset (if memory was cleared to factory default parameter values;
refer to section 3.2.4 on page 3-3)
Parameter Error (contact INFICON if this message appears)
Sensor Data Error (contact INFICON if this message appears)
System Data Error (contact INFICON if this message appears)
Searching followed by:
DeviceNet Detected (if a DeviceNet module is installed in the Control card)
DeviceNet Not Detected (if a DeviceNet module is not installed)
DeviceNet Error (if there is a problem with the DeviceNet module)
Initializing followed by:
Sensor n Detected (for installed Sensor cards)
Sensor n Not Present (for empty Sensor card slots)
Sensor n Initialization Failed (if there is a problem with a Sensor card)