3.3.2.6 Sensor Screen
On the Overview screen, press Sensor Data to display the Sensor screen (see
Figure 3-6). Data displayed on the Sensor screen is for an individual sensor. Data
includes Concentration (Conc), FACTORY ZERO status, Temp 1 and Temp 2.
Indicators (described in section 3.3.2.2 on page 3-6) display when appropriate.
Errors, Warnings or Errors & Warnings may be displayed, depending on
whether errors, warnings, or both are present. If no errors or warnings are present,
this area will remain blank.
Zero Reference, Reset Reference, Overview, Parameters, and Advanced Data
are available.
Figure 3-6 Sensor screen
During Calibration, pressing Zero Reference sets a new Zero Reference point.
Zero Reference is only available if the system meets the AT TEMP and STEADY
conditions and if ALLOW USER ZERO is set to YES on the Sensor Parameters
screen.
Reset Reference re-establishes the Factory Zero setting and is only available
when User Zero is displayed. Both buttons appear grey when they are
not available.
If Factory Zero is used, FACTORY ZERO is displayed.
If the sensor parameter ALLOW USER ZERO is set to NO, or if the Composer Elite
Sensor Control Unit is not connected to a Composer Elite Acoustic Sensor, (USER
ZERO NOT AVAILABLE) will be displayed next to FACTORY ZERO.
If a user-defined zero is being used, USER ZERO is displayed.