4.1.10 What Happens if the Reference Zero
is Improperly Set?
If the Reference Zero is set when the conditions are wrong, i.e., not completely
purged of precursor gas:
1 A permanent concentration offset is introduced equal to the residual precursor
concentration when the new Reference Zero is applied. This is not corrected
until a proper Reference Zero is performed.
2 The sensitivity is altered, leading to an error proportional to the concentration.
Figure 4-6 Impact of wrong zero frequency
NOTE: Factory Zero will be restored if one of the carrier gas parameters
(molecular weight or ) is altered or if TEMP 1 Setpoint is changed or if
Allow User Zero is changed from Yes to No.
TMIn in H2: Impact Of Wrong Zero Frequency
-0.005
-0.025
-0.015
-0.035
0.0 2.01.51.00.5
True Concentration (mole-%)
Error (mole-%)
Zeroed at 10 Hz below
true carrier frequency
Zeroed at 20 Hz below
true carrier frequency