User guide

5 - 13
PN 074-566-P1B
Composer Elite Operating Manual
5.4.2.2.2 Sensor 1-5 Data Set
All of the basic and advanced data associated with a single sensor. The data set
can be requested for any of the 5 possible sensors.
Table 5-13 Sensor 1-5 data set
Name Range Data Type Note
Cached Data Set 1-5 Byte Byte
Boards Present
Indicator
0 – 0x1F Byte Byte
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Table 5-18
Integer Status bits Refer to Table 5-18.
SS bit is not relevant in the data set format.
Measurement
Mode
IDLE = 0
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded Current System Measurement Mode.
Concentration 0.0 – 100.0
mole %
Float The current concentration value.
Temperature 1 30.0 – 65.0 C Float The current temperature value of first
temperature sensor.
Temperature 2 30.0 – 65.0 C Float The current temperature value of second
temperature sensor.
Errors ------ Integer Current error value. Refer to Table 5-22 for a
complete description of the errors.
Warnings ------ Integer Current warning value See Table 5-23 for a
complete description of the warnings.
Heater 1 Status 0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Heater 2 Status 0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sample Number 0 – 255 Byte The current sample number
Frequency Float
Amplitude Float
Heater 1 Current
Drawn
Float
Heater 2 Current
Drawn
Float
R Squared Float