User guide

5 - 14
PN 074-566-P1B
Composer Elite Operating Manual
5.4.2.2.3 All 5 Sensor Card Data Set
This data set returns a compound set of current data values for all five of the sensor
cards.
Standard Error Float
Q Float
PWM 1 Integer
PWM 2 Integer
Acoustic Gain Byte
D Gain Heater 1 Float
I Gain Heater 1 Float
P Gain Heater 1 Float
D Gain Heater 2 Float
I Gain Heater 2 Float
P Gain Heater 2 Float
RAV Float NOTE: RAV value is calculated during the
Track and Q Track measurement modes. The
RAV value will remain steady at the last RAV
value calculated during the other
measurement modes.
Controller Card
Board Temperature
Float
Sensor Card Board
Temperature
Float
Table 5-13 Sensor 1-5 data set (continued)
Name Range Data Type Note
Table 5-14 All sensor card data set
Name Range Data Type Note
Cached Data Set 6 Byte Numeric value of the data set cached.
Boards Present
Indicator
0 – 0x1F Byte Bitwise: Bits 0 – 4 set indicate sensor board 1
– 5 present, respectively.
Sensor 1
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Table 5-18
Integer Status bits Refer to Table 5-18.
SS bit is not relevant in the data set format.
Sensor 1
Measurement
Mode
IDLE = 0
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded Current System Measurement Mode.
Sensor 1
Concentration
0.0 – 100.0
mole %
Float The current concentration value.
Sensor 1
Temperature 1
30.0 – 65.0 C Float The current temperature value of first
temperature sensor.