User guide

5 - 17
PN 074-566-P1B
Composer Elite Operating Manual
Sensor 4
Heater 1 Status
0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 4
Heater 2 Status
0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 4
Sample Number
0 – 255 Byte The current sample number.
Sensor 5
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Table 5-18
Integer Status bits Refer to Table 5-18.
SS bit is not relevant in the data set format.
Sensor 5
Measurement
Mode
IDLE = 0,
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded Current System Measurement Mode.
Sensor 5
Concentration
0.0 – 100.0
mole %
Float The current concentration value.
Sensor 5
Temperature 1
30.0 – 65.0 C Float The current temperature value of first
temperature sensor.
Sensor 5
Temperature 2
30.0 – 65.0 C Float The current temperature value of second
temperature sensor.
Sensor 5
Heater 1 Status
0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 5
Heater 2 Status
0 - 6 Short 0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 5
Sample Number
0 – 255 Byte The current sample number.
Table 5-14 All sensor card data set (continued)
Name Range Data Type Note