User guide

5 - 32
PN 074-566-P1B
Composer Elite Operating Manual
1 Current
Temperature
Temperature Zone 1|2 30.0 -
65.00(ºC)
99.999
Float Current Zone Temperature 1 for
Acoustic Sensor's resonant chamber.
(99.999 if cable is disconnected.)
30.0 -
68.00(ºC)
99.999
Float Current Zone Temperature 2 for
Acoustic Sensor's inlet tube. (99.999 if
cable is disconnected.)
3 Heater Status Heater ID 1|2 0-6 Short Status of Acoustic Sensor's resonant
chamber heater (1) and inlet tube
heater (2):
0 = Off
1 = Not at Temperature
2 = At Temperature
3 = Failed
4 = Cable Disconnected
5 = Near Temperature
6= Over Temperature
7 Boards
Present
Active Sensor Boards.
Bits 0-4 indicate sensor
board 1-5 present.
0x1F Byte 0 = board not present.
1 = board present.
8 Keyboard Lock
State
0|1 Byte 0 = Locked, 1 =Unlocked.
9Errors/
Warnings
Code
Errors 0 - 0xFFFF Short Current error value. See Table 5-22.
Warnings 0 - 0xFFFF Short Current warning value. See Table
5-23.
Table 5-22 System and hardware errors
Error
Number
Description Error Code Set by
CC1
or SC2
Sensor
Specific
(SS3)
Set/Clear
Trigger
General Notes/Actions
32 Manual Parameter
Reset Required
0x8000 0000 CC Does not
clear
ACTION(CC): Parameters will
lock internally. Operation will be
suspended.
NOTE: Expected to occur
because of any errors that occur
even though the flash versions
are compatible.
31 Parameter Flash HW
Failure
0x4000 0000 CC Does not
clear
ACTION(CC): Parameters will
lock internally. Operation will be
suspended.
NOTE: Indicates that the flash is
not responding correctly to the
read/write requests.
30 Image Flash HW
Failure
0x2000 0000 CC Cleared on
read
NOTE: Indicates that the flash is
not responding correctly to the
read requests.
1
CC = Controller Card,
2
SC = Sensor Card,
3
SS = Specific Sensor
Table 5-21 Status commands (continued)
S
Command
ID
Name Parameter Range
(low/high)
Data
Type
Note