User guide

5 - 33
PN 074-566-P1B
Composer Elite Operating Manual
29 DeviceNet HW Failure 0x1000 0000 CC Self Clear NOTE: Indicates that the
DeviceNet module is
self-reporting a hardware failure.
28 Temperature HW
Failure
0x0800 0000 CC Self Clear NOTE: Indicates that the
temperature reading being
reported is invalid.
20 Temperature 2
Calculated Out of
Range
0x0008 0000 CC Self Clear ACTION(CC): Temperature 2
parameter value will be held at
the limit.
19 Carrier and Precursor
Gases Chosen Are
Identical
0x0004 0000 CC SS Self Clear
18 CC Has Lost
Communications with
SC
0x0002 0000 CC SS Self Clear
17 SC/CC Major Version
Mismatch
0x0001 0000 CC SS Self Clear ACTION(CC): Operation will be
suspended.
16 Sensor (Cell)/Cable
Not detected
0x0000 8000 SC SS Self Clear ACTION(SC): SC will remain in
Ready/Idle state until connected.
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
15 Heater 1 Temperature
- Outside Operating
Range
0x0000 4000 SC SS Self Clear ACTION(SC): SC disables
Heater
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
NOTE: Temperature outside
operating range.
14 Heater 2 Temperature
- Outside Operating
Range
0x0000 2000 SC SS Self Clear ACTION(SC): SC disables
Heater
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
NOTE: Temperature outside
operating range.
13 Heater 1 Current -
Expectations Not Met
0x0000 1000 SC SS Self Clear ACTION(SC): SC disables
Heater.
ACTION(CC): CC may
reallocate any power
management resources
assigned to the SC.
NOTE: Does not meet
expectation based on PWM
duty.
Table 5-22 System and hardware errors (continued)
Error
Number
Description Error Code Set by
CC1
or SC2
Sensor
Specific
(SS3)
Set/Clear
Trigger
General Notes/Actions
1
CC = Controller Card,
2
SC = Sensor Card,
3
SS = Specific Sensor