User Manual

7 - 7
IPN 074-289L
Composer Operating Manual
S(tatus)
00 Current Data float float float float byte byte
(frequency amplitude concentration/phase temperature sample-number undefined)
01 Current Temperature float
03 Digital Inputs byte
04 Relay Values byte
05 Configuration Switches short integer
06 Analog input (voltage) float
07 Error Log 8 bytes
See definition of error log below.
08 State of Keyboard lock byte (0=unlocked,1=locked)
09 Current data plus flow float float float float byte byte float float
(frequency amplitude concentration/phase temperature sample-number undefined bubbler
flow dilution flow) Note: if dilution flow not in use its value will be 0.
10 State Status bit pattern as follows: byte
System Mode Bit: 3 2 1 Remote Stop Tracking Command issued
Idle 0 0 0
Amp Track 0 0 1
Sweep 0 1 1
Tes t 1 0 0
Error Idle 1 0 1 Fatal Error present
Control State Bit: 6 5 4
Manual 0 0 0
Bypass 0 0 1
Hold 0 1 0
Control 0 1 1
Stop 1 0 0
11 Drive Level and Amp Range byte, byte (Drive (0,6,12), Range (1-5))
12 Control In float
13 Composer Configuration byte
14 Analog In Option float
15 Bubbler Flow Dilution Flow float, float
98 returns designer command measured
values set up with R99.
99 Returns designer command codes
set up with R99 (see see page 7-8)float
Input Response
Input # 1234567
Bit #6543210
Relay Response
Relay #54321
Bit #43210
0, 1, 2
Standard
Bulk Flow
SW#8 CFGR2 on
Additional Optional
Analog Input
SW#8 CFGR2 off
With Optional Analog Card