Owner manual

tira90e1 (2014-01) RS232C.cp 9
Parameter name Data
type
Access
Right
Byte
Type
Byte
No. 1
Byte
No. 2
Byte
No. 3
Comment
SWDateMonthDay uint16 R H-Byte 0x00 0xD6 <value> Software version date Month in Hex
e.g. 0x03 = March
L-Byte 0xD7 <value> Software version date Day in Hex
e.g. 0x19 = 19
Part No. uint8 R Byte 0 0x00 0xD8 <value> Part number as ASCII string
uint8 R Byte 1 0x00 0xD9 <value> (Max 20 byte)
uint8 R 0x00 <value> e.g. 378-000
uint8 R Byte 19 0x00 0xE8 <value> (Last digit: null terminator)
RemainingZero sint16 R H-Byte 0x00 0x48 <value> Max. remaining offset value
L-Byte 0x49 <value>
ExtendedError uint8 R H-Byte 0x00 0x36 Bit 0 PT1000 fault
H-Byte
Bit 1 Heater block overtemp.
Bit 2 Electronic overtemp.
Bit 3 Zero adjust error
Bit 4 Reserve
Bit 5 Reserve
Bit 6 Reserve
Bit 7 Reserve
ExtendedError uint8 R L-Byte 0x00 0x37 Bit 0 Atm. pressure out of range
L-Byte
Bit 1 Temperature out of range
Bit 2 Reserve
Bit 3 Reserve
Bit 4 Cal. mode wrong
Bit 5 Pressure underflow
Bit 6 Pressure overflow
Bit 7 Zero adjust warning
SensPressRange uint8 R 0x00 0x38 0 F.S.R. = E-3
(Exponent)
1 F.S.R. = E-2
2 F.S.R. = E-1
3 F.S.R. = E 0
4 F.S.R. = E+1
5 F.S.R. = E+2
6 F.S.R. = E+3
7 F.S.R. = E+4
SensFSR uint8 R 0x00 0x39 0 Mantissa = 1.0
(Mantissa)
1 Mantissa = 1.1
2 Mantissa = 2.0
3 Mantissa = 2.5
4 Mantissa = 5.0
FirmwareRevisionCPU2
uint8 R Byte 0 0x00 0x10 <value> Software version Cube application
program
uint8 R Byte 1 0x00 <value>
uint8 R x <value> e.g. 1.6.0.14
uint8 R Byte 19 x <value> (Last digit: null terminator)
ImageRevisionCPU2 uint8 R Byte 0 0xE0 0x24 <value> Image version of WinCE operating
system
uint8 R Byte 1 0xE0 <value>
uint8 R 0xE0 <value> e.g. WinCE60
uint8 R Byte 19 0xE0 <value> (Last digit: null terminator)
(continued)