User Manual

3 - 70
IPN 074-471-P1B
HAPSITE ER Operating Manual
3.9.1 Quick Reference SOP —
GC/MS Mode with HSS in Portable Mode
1 Verify that the desired sampling configuration is installed.
2 Insert the Transfer Line.
3 If the system is Shutdown or in Extended Standby, either power on the
HAPSITE or take the system out of Standby. If a Concentrator is installed, the
HAPSITE will begin preparing a Concentrator method. See Section 3.4.3,
Procedure for Running Concentrator Methods, on page 3-25.
3a If the SYSTEM IS NOT READY message is displayed, touch PREPARE
ANALYZE. Alternately, use the arrow keys to highlight PREPARE ANALYZE
and push OK SEL.
4 Insert the sample/blank vial into a well and an empty vial into another empty
slot. Place the needle into sample/ blank vial.
5 Touch RUN ANALYZE or push ANALYZE RUN when the SYSTEM IS READY
screen is displayed.
6 Remember: when the screen prompts, Collecting Sample Now For and
Sampling Is Done, no action is required from the user.
7 When prompted, put the needle into the empty vial and press OK to purge.
8 When the run is complete, a METHOD FINISHED prompt will appear.
9 Refer to Section 3.5.1, View Results/View Reports, on page 3-32 for
information on data review.
CAUTION
The Concentrator feature has increased sensitivity. Take
care to avoid saturating the HAPSITE.