User Manual
3 - 71
IPN 074-471-P1B
HAPSITE ER Operating Manual
3.10 SituProbe
For complete information on SituProbe methods, see Chapter 14, SituProbe. For
instructions on setting up the SituProbe, refer to Section 2.7, SituProbe, on page
2-24.
3.10.1 Procedure for SituProbe Operation
1 Verify that the desired sample configuration is installed (i.e., Concentrator).
2 Make sure that the Transfer Line is installed.
3 When powered on or taken out of Extended Standby, the HAPSITE will
automatically start preparing a SituProbe method. If the SituProbe method that
the HAPSITE begins preparing is not the desired one, refer to Section 3.2,
Selecting a Different Method Using the SELECT METHOD Icon, on page 3-13.
4 The HAPSITE will display the progress of its preparation. Refer to section
Chapter 3, Step 5 for more information.
5 If using the water vessel, fill the vessel to the top of the outlet connection. See
Figure 3-120.
Figure 3-120 SituProbe Vessel
6 Insert probe into vessel or water source.
6a If running a SituProbe method that requires a Concentrator, the HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to section 3.1.3 on page 3-11.
7 When the HAPSITE has finished preparing, a SYSTEM IS READY message
will appear with a prompt to press Survey or Analyze to begin sampling.
NOTE: A blank run is recommended before running a sample.
Top Of Outlet
Connection