7.4 AutoTune Failure
Occasionally, AutoTune may fail. Follow the procedures outlined below before
continuing.
1 When AutoTune fails, check Internal Standard canister pressure and expiration
date. Replace if necessary.
2 Rerun AutoTune by touching the HAPSITE icon on the front panel, touching the
HAPSITE System icon, touching the Tune icon and then, the Run AutoTune
button.
3 If AutoTune fails again, select File on the Laptop. From the drop down menu,
select View Tune Reports. See Figure 7-9.
Figure 7-9 Selecting View Tune Reports