User Manual

7 - 32
IPN 074-471-P1B
HAPSITE ER Operating Manual
3 After making Ion Energy adjustments:
3a If the red boxes no longer appear, indicating the Actual Resolution is within
range, proceed to Step 4. See Figure 7-36.
3b If the HAPSITE is failing to Tune, then repeat the Manual Tune procedure.
Proceed to section 7.6.1, Adjusting Base Peak Gain, on page 7-25. If this
procedure is repeated more than twice, contact INFICON.
Figure 7-36 Good Manual Tune
4 Save the Tune file. Click Save Tune button on the upper Toolbar. Refer to
Figure 7-36 and see Figure 7-37.
Figure 7-37 Save Tune Button
246 180 230
263 185 250
282 190 255
Table 7-1 Manual Adjustments Guidelines (continued)
MASS IE low IE high