Instruction Manual

Operation 41
* compatib. LDS1000/LDS2010 or binary protocol
611 W
Machine factor too high,
(machine factor)
81 153
1.00E+
04
Machine factor adjustment
inaccurate, split flow factor
too high
612 W
Machine factor lower
than last,
(machine factor)
81 153
< 50%
old
value
Split flow factor has
changed
613 W
Machine factor higher
than last,
(machine factor)
81 153
> 200%
old
value
Split flow factor has
changed
625 W
Int testleak not set,
(int. Testleak)
00
Leakage rate of internal
test leak is still set to
factory setting
626 W
Ext testleak not set,
(ext. Testleak)
00
Leakage rate of test leak is
still set to factory setting
630 W Calibration request 0 0
Temperature change of
5°C, speed was changed
since last calibration, 30-
minute switch-on time and
still no calibration
conducted
7xx temperature errors (pre-amplifier, electronics)
700 W
Preamplifier temp too
low,
(temperature [°C])
33 60 2°C Temperature too low
702 W
Preamplifier temp too
high,
(temperature [°C])
32 124 60°C Temperature too high
703 E
Preamplifier not
connected
33 60
Pre-amplifier defective,
cable not plugged in
710 W
Max. MSB temperature
exceeded,
(temperature [°C])
54 44 55°C Temperature too high
711 E
MSB temperature far
too high,
(temperature [°C])
54 44 65°C Temperature too high
8xx not used
9xx Maintenance warnings (e. g. TMP)
901 W
Maintenance bearing/
lubricant
99 99 3 years
TMP maintenance
necessary
Error
number
Warning/
Error
Error message
LDS3000
Error
number
LDS1000
protocol
Error
number
ASCII
protocol*
Limit
values
Cause