3-16 Operation of the Protec P3000
kina26e Chapter 3.fm Technical Handbook(1209)
If the value measured for the PRO-Check is out of range, the message "recalibration
required!" will be displayed in inverted colors.
For returning to the measuring mode, please press the right probe button or press
OK” on the main display.
For Protec P3000XL only
For the Protec P3000XL both calibration factors for
HIGH FLOW and LOW FLOW mode
will be verified (see fig. 3-16). However, only the information for the selected flow
mode will be relevant for passing or failing the verification. The results for the non-
selected mode will be displayed for reference at the bottom of the results screen.
Fig. 3-16 Results of proof function for Protec P3000
Fig. 3-17 Results of proof function for Protec P3000XL