User guide

Equipment Settings 4-26
kina26e Chapter 4.fm Technical Handbook(1209)
Printer manual
The
PRINTER MANUAL protocol allows to send leak rates on pressing the right sniffer
probe button.
Notice When in any of the two printer modes, no ASCII commands must be sent
to the Protec P3000 as this will interrupt the printer mode.
When set to I·Guide mode the PRINTER MANUAL mode is without function.
Printer auto
In
PRINTER AUTO the leak rate will be sent any time the set trigger level is exceeded.
At the time the signal drops below the trigger level the maximum detected leak rate
will be sent.
When set to I·Guide mode the leak rate results will be sent after each test point
together with the test point number. After the test cycle has been completed, the
summarized global leak rate will be sent.
Format
Date Time Gas equivalent Leak rate Unit
Example
03.05.2005 12:31 He 6x10
-4
mbar l/s
Format
Date Time Gas equivalent Leak rate Unit
Example
03.05.2005 12:31 R134a 2.34 g/a
Example:
Point no. Date Time Leak rate Unit
Point 01 8.3 E-6 mbar l/s
Point 02 2.0 E-6 mbar l/s
Point 03 8.4 E-6 mbar l/s
Point 04 1.2 E-6 mbar l/s
Global 13.04.2007 11:57:03 1.8 E-6 mbar l/s