User guide
Equipment Settings 4-28
kina26e Chapter 4.fm Technical Handbook(1209)
4.6 The Info Menu
The INFO MENU lists all internal data that may be useful for any trouble-shooting of the
Protec P3000. The
INFO MENU consists of 10 pages. On each page the upper left
button “
BACK” allows to leave the info menu, the lower left button allows to go back
one page whereas the lower right button allows to go to the next page. The page
number is displayed in the upper right corner.
Menu item Format Description
Page 1: General Data
Foreline pressure mbar
Flow sccm Flow through sniffer line
Time since power on Min
Operation hours h
Serial number 11-digit number
Wise Serial number
Software version x.xx.xx
Circuit temperature °C Temperature of main board
Test leak temperature °C
Sensor block Protec P3000 1 / 2 / 3
1 = Protec P3000
2 = Protec P3000 with
extended measurement
range
3 = Protec P3000XL
Page 2: Sensor Global
I akt. A
B-Filter A
I-Filter A
Q-Filter A
Leak rate mbar l/s helium leak rate
PWM-Membrane Integer
Heater Voltage V
High Voltage V
Wise-State Integer