Protec P3000 Messages 5-5
kina26e Chapter 5.fm Technical Handbook(1209)
W42 Pressure at calibration differs
between
HIGH and LOW FLOW
For Protec P3000XL only in
HIGH
FLOW mode:
Pressure at Wise Technology
sensor differs from LOW FLOW
mode to high flow mode (during
start-up or calibration only)
Filter in sniffer line blocked
Internal filter in main unit blocked
Capillary blocked
Replace sniffer tip filter and recalibrate
(See Section 7.4)
Replace internal filter* and recalibrate
Replace sniffer line and recalibrate!
Or: Replace probe cable!* and
recalibrate
See Section 7
W53 Flow at calibration out of limits! Only for Protec P3000 or Protec
P3000XL in LOW FLOW mode:
Flow at calibration higher than
upper flow limit or lower than lower
flow limit or limit low flow error (if
enabled, whichever comes first)
Filter in sniffer line blocked!
Internal filter in main unit blocked
Leak in (low flow) capillary
Wrong flow limit or limit low flow
error settings (for
LOW FLOW mode)
Replace sniffer tip filter and recalibrate!
(see section 7.4)
Replace internal filter* and recalibrate!
Replace sniffer line (or: replace probe
cable*) and recalibrate!
Check flow limit settings and limit low
flow error (for LOW FLOW mode)! See
section 4.4.1
W54 Flow at calibration out of limits! Only for Protec P3000XL in
HIGH
FLOW mode:
Flow at calibration higher than
upper flow limit
or lower than lower flow limit or limit
low flow error (if enabled,
whichever comes first)
Filter in sniffer line blocked
Internal filter in main unit blocked
Leak in probe cable
Wrong flow limit or limit low flow
error settings (for
HIGH FLOW mode)
Replace sniffer tip filter and recalibrate!
(see section 7.4)
Replace internal filter* and recalibrate!
Replace sniffer line (or: replace probe
cable*) and recalibrate!
Check flow limit settings and limit low
flow error (for HIGH FLOW mode)! See
section 4.4.1
Error
no.
Plain text message Possible Reason Recommendation