User guide

5-6 Protec P3000 Messages
kina26e Chapter 5.fm Technical Handbook(1209)
E55 Flow through sniffer below error
limit
Only for Protec P3000 or Protec
P3000XL in LOW FLOW mode:
Flow through sniffer line below limit
low flow error for LOW FLOW mode
Filter in sniffer line blocked!
Internal filter in main unit blocked
Limit low flow error for LOW FLOW
mode too high
Replace sniffer tip filter and recalibrate!
(see section 7.4)
Replace internal filter* and recalibrate!
Check limit low flow error (for LOW FLOW
mode)! See section 4.4.1
E56
HIGH FLOW below error limit Only for Protec P3000XL in HIGH
FLOW mode:
Flow through sniffer line below limit
low flow error for HIGH FLOW mode
or pressure at Wise Technology
sensor lower than corresponding
lower pressure limit
Filter in sniffer line blocked!
Internal filter in main unit blocked
Limit low flow error for HIGH FLOW
mode too high
Replace sniffer tip filter and recalibrate!
(see section 7.4)
Replace internal filter* and recalibrate!
Check limit low flow error (for HIGH FLOW
mode)! See section 4.4.1
W59 EEPROM parameter queue
overflow!
May occur if software “update” to
older version is performed
Restart the Protec P3000, if problem
persists, call nearest INFICON service
representative!
W60 All EEPROM parameter lost!
Please check your settings!
New EEPROM has been installed,
EEPROM on motherboard is virgin
All settings in software menu are reset
to default! Please enter your settings
again!
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty
Replace EEPROM*
W61 EEPROM parameter initialized! Software update performed and
new parameters have been
introduced
Newly introduced parameters are
listed below warning
Acknowledge warning
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty
Replace EEPROM*
W62 EEPROM parameter lost! Parameter has been modified
during software update and reset
to default value
Affected parameters are listed
below warning
Check setting of modified parameters in
corresponding software menu and set
to desired value!
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty
Replace EEPROM*
W64 There are outstanding
warnings!
Acknowledged but still valid
warnings will be repeated every 2
hrs or on each new power on
Please double-check the warnings!
Error
no.
Plain text message Possible Reason Recommendation