9.6.2 SpO2 Alarm Setup
Select SpO2 Menu Select Alarm Setup tab, as shown in Figure 9-3.
Figure 9-3: SpO2 Alarm Setup Menu
You can set the following alarm parameters for SpO2 module: SpO2,PR,PI, (PI setting is
enabled only when Masimo SpO2 module is used. Set the alarm according to your needs)
9.7 SpO2 Troubleshooting
This section lists the problems that might occur. If you encounter problems when using the
monitor or accessories, check the table below for troubleshooting. If the problem persists, contact
your service personnel.
The following message may appear in Pleth waveform area:
Problem Corrective Action
No SpO2 parameter window
and waveform area are
displayed
1
Check if SpO2 module is plugged properly;
2Check if the SpO2 parameter and waveform display is turned
ON in “View Layout”.
3
Make sure SpO2 module and Main Unit of the monitor are
connected.
“---” appears in SpO2 value
1
Check if the SpO2 sensor and adapter cable are connected
tightly. Replace the SpO2 sensor or cable if necessary.
2
Check "SpO2 sensor off" alarm occurs, reconnect the sensor.
Message Explanation
“Long searching time” SpO2 searching time is too long
“Searching pulse…” This monitor is searching the pulse signal
“SpO2 sensor off”
The SpO
2
sensor has become detached from the patient
or the module
“SpO2 communication error”
There is a communication problem with the SpO
2
measurement board