Checking the system with control solution Checking the system with control solution
Testing tips:
• Please make a test after reading the instructions for use of control solution
thoroughly.
• Keep clean of the test area on the meter.
• Keep test strips and control solutions in proper condition according to the
instructions for use provided, such as expiry date, storage condition and so
forth.
• Before testing, make sure that the test strips and the meter for lipid profile
are at the temperature of 18-30°C(64-86°F) and those for glucose at
10~40°C(50~104°F).
Caution
If your control solution test falls out of range, please follow the next steps
before contacting customer support:
• Check the expiry dates on all the products you are using.
• Try another control solution test.
• If this test falls out of range, try another control solution test with a new
unopened bottle of strips.
After following the appropriate steps and the control solution test still falls out
of range, do not perform a test. Please contact the infopia customer service
at +82-31-423-6170 or sales@infopia21.com.
• The control solution range is not the recommended range for your test.
• Discard the used control solution and the test strips carefully according to
the local regulations.
• Glucose test
Testing procedure
1. Shake the control solution before use.
3. Check the code number on the display of the meter with
that on the strip bottle.
4. Insert the test strip and press Mode ( ) button. Press
Up() button and the meter displays the symbol of control
solution( ).
5. In 5 seconds the symbol of sample blood is blinking.
6. Place a drop of solution on a clean, dry
surface (for example: the lid of your test
strip vial).
7. Apply the prepared control solution to the
test strip by using capillary rod.
8. Apply the control solution to the test area.
Caution : If you want to test all three test items (TC, HDL
and TG), please collect the control solution with
each capillary rod.
Caution : Whenever you change the test strip vial, Please
follow the code recognition procedure.
9. The test results will be displayed in about 2 minutes.