Operation Manual

24
© INNOLUX Corporaon │ X-Ray Products
outside the patient environment i.e. at least 1.5 m from the patient support or
shall be supplied via a separation transformer to reduce the leakage currents.
Any person who connects external equipment to signal inputs, signal outputs
or other connectors has formed a Medical Electrical System and is therefore
responsible for the system to comply with the requirements. If in doubt, contact
qualified medical technician or your local representative.
_____________________________________________________________
2.15 Disposal of Waste
It is illegal to place the device in the trash. Please contact the qualified local
recycling provides before disposing the device, The disposal procedure should
comply with local law.
2.16 Product Labels
The information of each label can be found in this section and the
corresponding location are also presented as the following figure. The
following labels are adhered to the INNOLUX Yushan X-Ray Flat Panel
Detector.