The transmitter is congured from a standard IBM compatible PC
by using the IPRO software (version 4.1 or later). IPRO 4.1 is com-
patible with Windows 98SE, Windows NT Workstation 4.0, Windows
2000 and Windows XP. When the transmitter is recongurated by a
PC no recalibration is needed.
Connect the IPRO communication cable between the serial port of
a PC (with IPRO software installed) and the PC connection port on
IPAQ-4L.
The IPAQ-4L must be powered, when you congure the
transmitter.
If IPRO (the IPAQ software) isn’t installed in your PC, install the
software (follow the instructions in ”IPRO”). Conguration from
PC is “on-line”, that is, the transmitter can be installed in the process
with input connected during conguration. The output is frozen while
transmission from PC to IPAQ take place. When transfer is done the
transmitter uses the new parameters.
NOTE: The label on the transmitter indicate allowed power supply
to the transmitter. By connecting false power supply the
transmitter may be damaged.
INTRODUCTION
IPAQ-4L is a fully universal and programmable 4-wire (separately
powered) transmitter. The high isolation provides safe and problem
free measurements. Offering both current and voltage output,
IPAQ-4L meets any process requirements. Designed for wall and
DIN-rail mounting.
IPAQ-4L is intended to be used in industrial environment.
DIMENSIONS
Universal
High-Isolation
4-wire Transmitter
CONFIGURATION
SWEDEN
GERMANY
USA
FINLAND
Since IPAQ-4L
is a precision instrument it´s recommended to check
the calibration once a year. Use equipment with an
inaccuracy of at least 5 times better than the checked one, e.g. if you
are checking a transmitter with an inaccuracy of 0.1% use instrument
with an inaccuracy of maximum 0.02%.
You can either perform a full recalibration of the transmitters stored
calibration parameters (Basic Calibration) or make a System
Error Correction (recommended). In each case use the IPRO software
and follow the instructions on the screen. When performing a System
Error Correction, please note that this is only a compensation for
errors at the measuring range in use.
CALIBRATION
86BIQ00007 2012-10
mm/inches
This product should not be mixed with other kind of scrap, after usage. It should be handled as an electronic/electric device.
PC connection
ADDRESS 7 Dearborn Road,
Peabody, MA 01960
USA
PHONE +1-978-826 6900
FAX +1-978-535 3882
E-MAIL inor-info@krohne.com
ADDRESS PO Box 9125
SE-200 39 Malmö
Sweden
PHONE +46-40-31 25 60
FAX +46-40-31 25 70
E-MAIL support@inor.se
ADDRESS Am Bocksborn 5
D-63571 Gelnhausen
Germany
PHONE +49-6051 14807
FAX +49-6051 14806
E-MAIL support.de@inor.se
ADDRESS Unikkotie 13
FI-01300 Vantaa
Finland
PHONE +358-10-421 7900
FAX +358-10-421 7901
E-MAIL support.@inor.se
INOR Process AB
INOR Transmitter GmbH
INOR Transmitter Oy
KROHNE Temperature
Division INOR

Summary of content (2 pages)