Instruction Manual

CONN ECTIONS
Item Part No.
IPAQ-L 70IPL00001
IPAQ-LX (ATEX) 70IPLX0001
IPAQ-LX (FM, CSA) 70IPLX1001
IPAQ-L(mA) Current input only 70IPL00003
Software and cable
Conguration kit IPRO with cables 70CFG00092
LIMI T ED WARRAN TY
INOR Process AB, or any other afliated company within the Inor Group (her-
einafter jointly referred to asInor), hereby warrants that the Product will be
free from defects in materials or workmanship for a period of ve (5) years from
the date of delivery (”Limited Warranty). This Limited Warranty is limited to
repair or replacement at Inor’s option and is effective only for the rst end-user
of the Product. Upon receipt of a warranty claim, Inor shall respond within a
reasonable time period as to its decision concerning:
1 Whether Inor acknowledges its responsibility for any asserted defect in
materials or workmanship; and, if so,
2 the appropriate cause of action to be taken (i.e. whether a
defective product should be replaced or repaired by Inor).
This Limited Warranty applies only if the Product:
1 is installed according to the instructions furnished by Inor;
2 is connected to a proper power supply;
3 is not misused or abused; and
4 there is no evidence of tampering, mishandling, neglect, accidenta da-
mage, modication or repair without the approval of Inor or damage
done to the Product by anyone other than Inor.
This Limited Warranty is provided by Inor and contains the only express war-
ranty provided.
INOR SPECIFICALLY DISCLAIMS ANY EXPRESS WARRANTY NOT
PROVIDED HEREIN AND ANY IMPLIED WARRANTY, GUARANTEE
OR REPRESENTATION AS TO SUITABILITY FOR ANY PARTICULAR
PURPOSE, PERFORMANCE, QUALITY AND ABSENCE OF ANY HIDDEN
DEFECTS, AND ANY REMEDY FOR BREACH OF CONTRACT, WHICH BUT
FOR THIS PROVISION, MIGHT ARISE BY IMPLICATION, OPERATION
OF LAW, CUSTOM OF TRADE OR COURSE OF DEALING, INCLUDING
IMPLIED WARRANTIES OF MER-CHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. EXCEPT AS PROVIDED HEREIN,
INOR FURTHER DISCLAIMS ANY RESPONSIBILITY FOR LOSSES,
EXPENSES, INCONVENIENCES, SPECIAL, DIRECT, SECONDARY OR
CONSEQUENTIAL DAMAGES ARISING FROM OWNERSHIP OR USE
OF THE PRODUCT.
Products that are covered by the Limited Warranty will either be repaired or
replaced at the option of Inor. Customer pays freight to Inor, and Inor will pay
the return freight by post or othernormalway of transport. If any other type
of return freight is requested, customer pays the whole return cost.
DATA (shortfor m)
ORDERING TABLE
IPAQ-LX EX-DATA
Approval Demko 03 ATEX 134461X
CE 0539 II (1)G [Ex ia] IIC
Control Drawing 88DRW00022
IPAQ-L/LX/L(mA) are designed to t on a standard DIN Rail.
IPAQ-LX must be installed in safe area.
Connect input, output and power supply acc. to g. 1-10.
INST A LLATION
Power supply: IPAQ-L, IPAQ-L(mA) 7.5 to 36 VDC
IPAQ-LX 8 to 30 VDC
Isolation in/out: 1500 VAC
Output: 4-20 mA
Operating temperature: -20 to +70 °C
T
1
T
2
6
+ -
5
mV
RTD
(Pt100)
T/C
RTD
100%
0%
RTD
RTD
(Pt100)
100%
0%
RTD
4-wire connection
Pt100
Diff temperature T
1
> T
2
Potentiometer
3-wire connection
Potentiometer
4-wire connection
Voltage
mV
Thermocouple
Supply voltage U (VDC)
RTD
3-wire connection
Fig 3
Fig 4
Fig 5
Fig 6
Fig 7
Fig 8
Fig 9
IPAQ-L, IPAQ-L(mA) IPAQ-LX
!
!
Fig 1
Fig 2
IPAQ-L(mA)
R
LOAD
=(U-7.5)/0.22 IPAQ-L, IPAQ-L(mA)
R
LOAD
=(U-8)/0.22 IPAQ-LX
Fig 10
Current
mA
Approval FM, J.I. 0D6A8.AX, CSA 2007 Certicate 1863602
Class I, II and III, Division 1, Group A, B, C, D, E, F and G
Control Drawing 3-7852
Output (terminals 5-6)
(current loop)
U
i
: 30 VDC
I
i
: 100 mA
P
i
: 0,9 W
L
i
: 0 mH
C
i
: 1 nF
Input (terminal 1-2-3-4)
(Intrinsically safe sensor terminals)
U
o
: 30 V
I
o
: 27 mA
L
o
: 50 mH
C
o
:
52 nF
The equipment must be electrically connected (terminal 5 and 6)
via a certied isolating interface/zener barrier -and shall be
placed outside the hazardous area.
1)
Ri ≤ 10 Ω
1)