Programmer's Manual

Table Of Contents
Programmer Model 2740 English 89
Generator Reset
A generator reset can occur in response to a low generator battery or severe electromagnetic disturbance. The
generator serial number is cleared by the reset and must be re-entered during the next programming session.
When prompted by the generator reset screen,
1. Enter the numeric portion of the serial number which can be found in the patient’s medical records.
2. Select the Check button.
a. If the patient information displayed is correct, select the Configure button to configure the generator
with the entered serial number.
b. If the patient information is incorrect, enter the correct serial number and select the Check button.
Confirm that patient information is correct and select the Configure button.
Note: The generator will reset to default parameters.
3. If the serial number is unavailable, select the Cancel button. When prompted, select the generator model
and select the Configure button. The generator will be configured with a temporary serial number and the
session report will be stored under the temporary serial number without a patient name or ID.
Exporting
Cannot Export Report
If the Export button is disabled, select the Enable USB Report Export check box on the Programmer Settings screen.
Export Failure Screen
If an export failure screen displays, follow these steps:
1. Disconnect the Inspire USB flash drive from the tablet.
2. Reconnect the Inspire USB flash drive to the tablet.
3. Wait 30 seconds and retry export.
4. If these measures do not resolve the problem, use a different USB drive with FAT32 formatting.
Missing Information
If an exported report is missing patient information, screenshots, or advanced settings such as electrode
configuration or level measurement settings, select the check box that corresponds to the desired information on
the Export Report screen and repeat the export.
Note: The option to include patient information, screenshots, and advanced settings in an exported report always
displays regardless of whether the content is available for a particular report. Review the report on screen to
determine if the content is available.